This free SPI study guide teaches to the ARDMS Sonography Principles & Instrumentation exam — every content domain the American Registry for Diagnostic Medical Sonography tests, organized the way the official outline is built.[1] The SPI is the physics-and-instrumentation examination required for all ARDMS credentials; passing it plus a specialty exam (Abdomen, OB/GYN, Echocardiography, and others) earns a registered credential such as RDMS or RDCS.
This guide is interactive, not a wall of text: every domain has a built-in checkpoint quiz, hover-able glossary terms, worked physics examples, labeled diagrams, and concept questions, so you learn by doing. Because medical imaging is a patient-safety field, the bioeffects and safety facts here — mechanical index, thermal index, ALARA, intensity, and the Nyquist limit — are verified against ARDMS and AIUM primary sources.[4]
Read this guide domain by domain, test yourself at each checkpoint, then round out your free SPI prep with our practice questions and flashcards.
SPI Exam Snapshot
| Detail | ARDMS SPI Exam |
|---|---|
| Questions | ~110 multiple-choice questions |
| Total time | 2 hours (includes a ~3-minute survey) |
| Score scale | Scaled 300–700; a scaled score of 555 passes (Pass/Fail) |
| Domains | 5 content domains — Doppler is the largest at 34% |
| Credential role | Required physics exam for all ARDMS specialties (RDMS, RDCS, RVT) |
| Format | Computer-based at a Pearson VUE test center |
| Scoring note | Scaled score is not a percentage and is not curved |
| Issuing body | American Registry for Diagnostic Medical Sonography (ARDMS) |
ARDMS reports the SPI domains and their official weights as approximate shares of the ~110 questions.[2] Spend your study time in proportion to the weights — Apply Doppler Concepts (34%) and Optimize Sonographic Images (26%) together are about 60% of the exam:
This guide teaches all five domains as five study modules, in the order they build on one another: the physics fundamentals first, then transducers, image optimization, Doppler, and finally clinical safety and quality assurance.
1 · Perform Ultrasound Examinations
About 23% of the exam (~25 questions). This domain holds the physics fundamentals — how sound waves propagate, interact with tissue, and form echoes — plus patient care, contrast, extended imaging, and artifact correction.[2]
Diagnostic ultrasound assumes a constant soft-tissue speed of 1,540 m/sto convert an echo’s return time into a depth.
- 1 · Transmit a short pulseThe transducer's piezoelectric element converts a voltage spike into a brief pulse of sound that travels into the body.
- 2 · The pulse reflects at interfacesAt each boundary where acoustic impedance changes, part of the sound reflects back toward the transducer as an echo.
- 3 · Time the round tripThe system measures how long each echo takes to return. Listening time, not transmit time, dominates the cycle.
- 4 · Compute depth (range equation)Depth = (1,540 m/s × round-trip time) ÷ 2. The ÷ 2 accounts for the down-and-back path. Echoes are placed on the image by depth.
A wrong assumed speed (a speed-error artifact) misplaces the echo’s depth.
Sound Waves & Propagation
Ultrasound is a mechanical pressure wave above 20 kHz; diagnostic imaging uses about 2–15 MHz. The key relationship is = ÷ :
In soft tissue the average propagation speed is 1,540 m/s, so . A 5 MHz beam has a wavelength of about 0.31 mm. Frequency, period, wavelength, and propagation speed are all source/medium properties — the operator cannot change frequency with a depth or gain knob.
Reflection, Attenuation & Impedance
(Z) = density × propagation speed. A reflection occurs wherever two tissues differ in impedance; the bigger the , the stronger the echo. — the progressive weakening of the beam from reflection, scattering, and absorption — increases with both depth and frequency:
| Interaction | What happens |
|---|---|
| Reflection | Echo returns at an impedance boundary; strongest when the beam hits perpendicularly |
| Refraction (Snell's law) | Transmitted beam bends at an oblique interface with different speeds |
| Scattering | Sound redirected in many directions by small/rough reflectors (e.g., red cells) |
| Absorption | Acoustic energy converted to heat — the main cause of attenuation |
| Attenuation (soft tissue) | About 0.5 dB per cm per MHz; total ≈ 0.5 × frequency × path length |
A (large, smooth — e.g., the diaphragm) returns the strongest echo when struck perpendicularly, while a diffuse (scatter) reflector returns weaker, angle-independent echoes. governs how the transmitted beam refracts at an oblique boundary.
Pulse-Echo & the Range Equation
Imaging works by pulse-echo: the system sends a brief pulse, times the echo’s round trip, and computes depth with the range equation:
The ÷ 2 accounts for the down-and-back path. depends on the (cycles × wavelength) and equals half of it — so shorter pulses (higher frequency, more damping) resolve structures lying along the beam better.
Patient Care, Contrast & Artifacts
This domain also covers the start-to-finish exam: verify patient identity and the appropriateness of the order, review clinical history and prior imaging, scan ergonomically to protect yourself from injury, and document representative images. Contrast agentsare gas-filled microbubbles that boost the reflectivity of blood (imaged at low MI so the bubbles aren’t destroyed). Panoramic (extended field-of-view), 3D (a rendered volume), and 4D (real-time 3D) extend what a single image shows.
Gray-scale artifacts you can identify and correct include (equally spaced parallel lines), mirror-image, , and — covered in depth in the image-optimization module.
Checkpoint · Domain 1 · Perform Ultrasound Examinations
Question 1 of 10
What is the primary effect of increasing the frequency of an ultrasound wave on tissue penetration?
2 · Manage Ultrasound Transducers
About 7% of the exam (~8 questions) — the smallest domain, but high-yield because the concepts recur everywhere. How a transducer makes and shapes the beam, how frequency is chosen, and how arrays steer and focus.[2]
Transducer Components
The heart of a transducer is the element, which converts voltage to vibration and back. Three more components shape the pulse:
| Component | Function |
|---|---|
| Active element (PZT crystal) | Converts voltage ↔ sound; thickness sets the operating frequency |
| Backing (damping) material | Shortens the pulse → better axial resolution, wider bandwidth, lower Q |
| Matching layer | Reduces the impedance mismatch with skin so more sound enters the body |
| Acoustic lens | Focuses the beam in the elevational (slice-thickness) plane |
Frequency & Transducer Selection
The element’s thickness sets the operating (resonant) frequency: thinner elements resonate higher. Frequency drives the central imaging trade-off:
- ✓ Short wavelength → better axial resolution
- ✓ Narrow beam → better lateral resolution
- ✗ Attenuates fast → shallow penetration
- Use for: vascular, thyroid, small parts
- ✓ Attenuates slowly → deep penetration
- ✗ Long wavelength → coarser axial resolution
- ✗ Wider beam → coarser lateral resolution
- Use for: abdominal, obstetric, cardiac
Rule: choose the highest frequency that still reaches the structure of interest. Wavelength (mm) ≈ 1.54 ÷ frequency (MHz) in soft tissue.
Imaging transducers are broadband (low Q-factor): heavy damping shortens the pulse for good axial resolution but widens the bandwidth. A dedicated probe is the opposite — a narrow-band, high-Q transducer transmitting a pure tone.
Array Types & Beam Formation
Modern transducers are arrays — rows of small elements fired with timed delays to steer and focus the beam electronically. The near zone narrows toward the focus; the far zone diverges (a larger aperture and higher frequency reduce divergence).
| Array | Image & typical use |
|---|---|
| Linear (sequential) | Rectangular field; high-frequency, superficial — vascular, small parts |
| Curvilinear (convex) | Wide sector-like field; lower frequency — abdominal, obstetric |
| Phased | Small footprint, electronically steered sector — cardiac (between ribs) |
| Annular | Symmetric focus in both planes; steered mechanically |
Checkpoint · Domain 2 · Manage Ultrasound Transducers
Question 1 of 10
In ultrasound physics, what describes the piezoelectric effect?
3 · Optimize Sonographic Images
About 26% of the exam (~29 questions) — the second-largest domain. Resolution, brightness controls, processing techniques, the various display modes, and the artifacts that mislead interpretation.[2]
Axial, Lateral, Elevational & Temporal Resolution
There are four kinds of resolution, and the exam tests how each is improved:
Best with short pulses (higher frequency, more damping)
Best at the focal zone, where the beam is narrowest
Axial resolution is always finer than lateral resolution in clinical imaging.
| Resolution | Set by / improved by |
|---|---|
| Axial (along the beam) | Spatial pulse length ÷ 2 — shorter pulse, higher frequency, more damping |
| Lateral (across the beam) | Beam width — best at the focal zone; place focus at the region of interest |
| Elevational (slice thickness) | Beam thickness — acoustic lens or a 1.5D array |
| Temporal (in time) | Frame rate — shallower depth, narrower sector, fewer focal zones, lower line density |
Gain, TGC, Dynamic Range & Output
Distinguish the controls that change brightness from the one that changes patient exposure:
| Control | What it does |
|---|---|
| Overall (receiver) gain | Amplifies all returning echoes — brightens signal AND noise; no patient exposure change |
| Time gain compensation (TGC) | Boosts deep echoes to offset attenuation → uniform brightness by depth |
| Dynamic range (compression) | Narrow = high contrast (fewer grays); wide = smoother, more grays |
| Output power | Changes the actual acoustic energy into the patient — affects bioeffects (TI/MI) |
Harmonic, Compounding, M-mode & Modes
forms the picture from harmonic frequencies generated within tissue, reducing clutter and artifact. Spatial compounding combines frames from several angles to cut speckle (at lower frame rate), and frequency compounding combines different frequency bands.
| Mode | Display |
|---|---|
| A-mode | Amplitude vs. depth along one line (the earliest format) |
| B-mode | Brightness-mapped 2D gray-scale image (the standard view) |
| M-mode | Depth of a single line over time — very high temporal resolution (e.g., heart valves) |
Image Artifacts
Artifacts are display errors you must recognize so you don’t mistake them for anatomy — and some are diagnostically useful:
| Artifact | Appearance & cause |
|---|---|
| Reverberation | Equally spaced parallel lines — sound bouncing between two strong reflectors |
| Comet tail / ring-down | Short bright trail / band deep to a small strong reflector or gas |
| Mirror image | Duplicate placed deeper than a strong reflector (e.g., diaphragm) |
| Shadowing | Dark band deep to bone/stone — high attenuation/reflection (suggests solid/calcified) |
| Enhancement | Bright band deep to a cyst — less attenuation (suggests fluid) |
| Side / grating lobe | Off-axis energy places an echo in the wrong lateral location |
| Speed error | Misplaced/split reflector — tissue speed differs from the assumed 1,540 m/s |
| Anisotropy | Tendon/nerve looks bright at 90° and dark at other angles |
Checkpoint · Domain 3 · Optimize Sonographic Images
Question 1 of 10
What role does the "time-gain compensation" 'TGC' play in ultrasound imaging?
4 · Apply Doppler Concepts
About 34% of the exam (~37 questions) — the single largest domain, so this is where to invest the most study time. The Doppler effect, the modes, aliasing and the Nyquist limit, and spectral-waveform interpretation.[2]
Doppler Effect, Shift & Angle
The is the change in reflected frequency when a reflector (red blood cells) moves relative to the transducer. The is proportional to the velocity of flow and to the cosine of the :
Accurate velocity
Large velocity errors
cos 90° = 0 → no shift
Detected shift ∝ cos θ. Keep the angle at or below 60°; a beam perpendicular to flow records no flow.
Because velocity depends on , a beam perpendicular to flow () detects no shift, and errors grow rapidly above 60°. Keep the angle at or below 60° and use angle correction.
CW, PW, Color & Power Doppler
| Mode | Key property |
|---|---|
| Continuous wave (CW) | Separate transmit/receive elements; measures very high velocities, NO aliasing, but no depth resolution |
| Pulsed wave (PW) | Samples a chosen depth (range resolution); limited by the Nyquist limit, so it CAN alias |
| Color Doppler | Mean velocity + direction as color on the 2D image; can alias |
| Power Doppler | Signal strength (amplitude); more sensitive to slow flow, NO direction, does not alias |
Aliasing, Nyquist & Wall Filter
is the classic pulsed-Doppler artifact: when the Doppler shift exceeds the (PRF ÷ 2), high velocities wrap around and display in the wrong direction.
Four ways to fix aliasing
- Raise the PRF / velocity scaleA higher PRF raises the Nyquist limit (PRF ÷ 2) so faster flow fits.
- Shift the baselineReassign more of the velocity scale to the dominant flow direction.
- Lower the transducer frequencyA lower frequency makes a smaller shift for the same velocity.
- Increase the Doppler angle (toward 60°)A larger angle lowers the measured shift — use with care.
The (a high-pass filter) removes low-frequency, high-amplitude signals from slow-moving vessel walls — but if set too high it erases genuine low-velocity venous or diastolic flow.
Hemodynamics & Spectral Waveforms
By the continuity equation, flow speeds up through a narrowing, so a high-grade stenosis produces a high-velocity jet and post-stenotic turbulence (filling in the spectral window = spectral broadening). The simplified Bernoulli equation estimates the pressure gradient across a stenosis:
Checkpoint · Domain 4 · Apply Doppler Concepts
Question 1 of 10
What parameter does the Doppler frequency shift primarily depend on?
5 · Provide Clinical Safety & Quality Assurance
About 10% of the exam (~11 questions). Bioeffects and the on-screen safety indices, infection control, and equipment quality assurance. This is core patient-safety content — get it exactly right.[4]
Bioeffects, MI, TI & ALARA
Diagnostic ultrasound has two established potential-harm mechanisms, each with its own on-screen index:
- Index: Thermal Index (TI)
- TI = power used ÷ power to raise tissue 1 °C
- Variants: TIS (soft tissue), TIB (bone), TIC (cranial)
- Worst at bone; worsened by long dwell time and Doppler
- Index: Mechanical Index (MI)
- MI = peak rarefactional pressure ÷ √frequency
- Driven by negative (rarefactional) pressure
- Higher MI → greater cavitation potential
The estimates non-thermal (cavitation) risk and equals the peak rarefactional pressure divided by the square root of frequency:
The estimates heating — the ratio of power used to the power needed to raise tissue 1 °C, with variants TIS (soft tissue), TIB (bone in focus), and TIC (cranial). The intensity descriptor most tied to heating is intensity. The shows TI and MI in real time so the operator can practice .[3]
Infection Control & Disinfection
Hand hygiene before and after every patient is the single most effective infection-control measure. Transducer reprocessing depends on what the probe contacts (the Spaulding classification):
| Probe contact | Required reprocessing |
|---|---|
| Intact skin only | Low-level cleaning + disinfection between patients |
| Mucous membranes / non-intact skin (endocavitary) | High-level disinfection between patients |
| Any probe | Inspect housing/cable for damage; remove from service if compromised |
Quality Assurance & Statistics
Routine QA uses a with known targets to test the system over time. Test parameters include axial/lateral resolution, the dead zone, vertical (depth) and horizontal (lateral) distance accuracy, uniformity, and low-contrast (lesion) detectability. Document results to detect equipment drift.
The outline also expects basic statistical concepts: sensitivity (the proportion of true positives correctly identified) and specificity (the proportion of true negatives correctly identified).
Checkpoint · Domain 5 · Clinical Safety & Quality Assurance
Question 1 of 10
Which of the following best describes the primary purpose of the ALARA principle in sonography?
How to Use This Study Guide
A study guide is a map, not the whole territory — use it alongside hands-on scanning and our free tools. Because the SPI is dense with formulas and definitions, spaced active recall beats one long cram: read a domain, take its checkpoint, then drill the gaps with flashcards and practice questions. Weight your time toward Doppler (34%) and image optimization (26%), the two largest domains, then shore up the physics fundamentals that underlie everything.
- 1
Read a domain here
Work through one domain at a time — physics, transducers, optimization, Doppler, then safety/QA.
- 2
Take the checkpoint
The quick check at the end of each domain exposes what didn't stick.
- 3
Drill the gaps
Send your weak domain straight into the free practice questions and flashcards.
- 4
Test full and timed
Sit full, timed practice to build pacing for ~110 questions in two hours, then review every miss.
SPI Concept Questions
Common SPI physics, instrumentation, and safety concepts the exam actually measures — at least one per content domain. Tap any card for a short, exam-ready answer backed by an official source (ARDMS or AIUM), then test yourself on them as flashcards.
SPI Glossary
Quick definitions for the terms you’ll see most across the ARDMS SPI exam:
- Acoustic enhancement
- Increased brightness deep to a weakly attenuating structure such as a fluid-filled cyst, because the beam there is less attenuated.
- Acoustic impedance
- A tissue's resistance to the propagation of sound, equal to density × propagation speed. Differences in impedance at a boundary create reflections (echoes).
- Acoustic impedance mismatch
- The difference in acoustic impedance between two adjacent media; the larger the mismatch, the stronger the reflection at the boundary.
- Acoustic shadowing
- A dark band deep to a highly attenuating or reflecting structure (bone, stone, calcification) where little sound reaches the tissue beyond.
- ALARA
- As Low As Reasonably Achievable — use the lowest acoustic output power and shortest scan time that still produce a diagnostic image, to minimize potential bioeffects.
- Aliasing
- A pulsed-Doppler artifact when the Doppler shift exceeds the Nyquist limit; high velocities wrap around and display in the wrong direction.
- Attenuation
- The progressive weakening of the beam (reflection, scattering, and absorption) as it travels. In soft tissue ≈ 0.5 dB per cm per MHz; greater at higher frequency and depth.
- Axial resolution
- The ability to resolve two structures along the beam (front to back); equal to half the spatial pulse length. Improved by higher frequency and shorter pulses.
- Cavitation
- A non-thermal bioeffect: the formation, oscillation, and violent collapse of gas bubbles driven by the peak rarefactional (negative) pressure; estimated by the MI.
- Color Doppler
- Overlays mean velocity and flow direction as color on the 2D image; encodes direction (toward/away) and is subject to aliasing.
- Continuous wave Doppler
- Doppler using separate transmit and receive elements running continuously: no aliasing and no velocity limit, but no depth (range) resolution.
- Doppler angle
- The angle between the Doppler beam and the direction of flow. Velocity accuracy depends on its cosine; keep it at or below 60 degrees.
- Doppler effect
- The change in the frequency of reflected sound when a reflector moves relative to the transducer — the basis for measuring blood-flow velocity.
- Doppler shift
- The difference between transmitted and received frequency caused by moving reflectors. Flow toward the transducer raises the frequency; flow away lowers it.
- Dynamic range
- The ratio (in dB) of the largest to smallest echo a system displays. Narrow range = high contrast (fewer grays); wide range = more shades of gray.
- Frequency
- The number of cycles per second (Hz). Diagnostic ultrasound uses about 2–15 MHz. Determined by the sound source, not by operator depth or gain controls.
- Gain
- Receiver amplification applied to all returning echoes; brightens the whole image (signal and noise) without changing acoustic output to the patient.
- Harmonic imaging
- Imaging formed from the higher harmonic frequencies generated within tissue (nonlinear propagation), giving fewer artifacts and better contrast.
- High-level disinfection
- The reprocessing required for transducers contacting mucous membranes or non-intact skin (e.g., endocavitary probes), per CDC guidelines.
- Lateral resolution
- The ability to resolve two structures side by side, across the beam; determined by beam width and best at the focal zone.
- Mechanical index
- An on-screen estimate of non-thermal (cavitation) bioeffect risk: MI = peak rarefactional pressure ÷ √frequency. Higher MI means greater cavitation potential.
- Nyquist limit
- The maximum Doppler shift a pulsed system can display without aliasing — equal to one-half the pulse repetition frequency (PRF ÷ 2).
- Output display standard
- The FDA/AIUM/NEMA standard (ODS) requiring the TI and MI to be displayed on screen so the operator can monitor and minimize potential bioeffects.
- Piezoelectric effect
- The transducer's conversion of voltage into mechanical vibration (transmit) and returning sound pressure back into voltage (receive).
- Power Doppler
- Encodes the strength (amplitude) of the Doppler signal, not velocity or direction; more sensitive to slow flow and does not alias.
- Propagation speed
- The speed at which sound travels through a medium. Diagnostic ultrasound assumes an average soft-tissue speed of 1,540 m/s (1.54 mm/μs).
- Pulse repetition frequency
- The number of pulses sent per second (PRF). Higher PRF raises the Nyquist limit (less aliasing) but limits maximum imaging depth.
- Pulsed wave Doppler
- Doppler that samples flow from a chosen depth (sample volume), giving range resolution but limited by the Nyquist limit, so it can alias.
- Reverberation artifact
- Multiple equally spaced echoes from sound bouncing between two strong reflectors, appearing as parallel lines descending into the image.
- Snell's law
- Describes refraction — the bending of the transmitted beam at an oblique interface between two media of different propagation speeds.
- Spatial pulse length
- The distance a single pulse occupies in tissue = number of cycles × wavelength. A shorter pulse gives better axial resolution.
- SPTA intensity
- Spatial peak temporal average intensity — the intensity at the spatial peak averaged over time; the parameter most relevant to thermal bioeffects.
- Thermal index
- An on-screen estimate of potential tissue heating: the ratio of acoustic power used to the power needed to raise tissue temperature by 1 °C (TIS, TIB, TIC).
- Time gain compensation
- An operator control (TGC) that amplifies echoes from deeper structures to offset attenuation, giving uniform brightness from top to bottom.
- Tissue-mimicking phantom
- A quality-assurance device with known targets used to test resolution, distance accuracy, dead zone, and low-contrast detectability.
- Wall filter
- A high-pass filter that removes low-frequency signals from slow-moving vessel walls. Set too high, it can erase true low-velocity (venous/diastolic) flow.
- Wavelength
- The length of one cycle of the wave. Wavelength = propagation speed ÷ frequency; in soft tissue ≈ 1.54 ÷ frequency (MHz). Higher frequency means shorter wavelength.
Free SPI Study Materials & Resources
Everything you need to prepare for the SPI exam is free here — no paywall, no sign-up. This guide is the foundation; pair it with the rest of our free SPI study materials for active recall, timed practice, and last-minute review:
- SPI Practice Test — exam-style questions across all five ARDMS domains, with explanations.
- SPI Flashcards — active-recall decks for the physics formulas, Doppler concepts, and safety facts.
SPI Study Guide FAQ
The Sonography Principles & Instrumentation (SPI) exam has approximately 110 multiple-choice questions. The exam is administered across five content domains: Perform Ultrasound Examinations (23%), Manage Ultrasound Transducers (7%), Optimize Sonographic Images (26%), Apply Doppler Concepts (34%), and Provide Clinical Safety & Quality Assurance (10%).
The SPI exam is two hours long, which includes a brief survey of about three minutes. With roughly 110 questions in two hours, you have just over one minute per question on average, so pacing and quick recall of physics formulas matter.
ARDMS uses a scaled score of 555 to pass all of its examinations, on a scale that ranges from 300 to 700. The scaled score is not a percentage of correct answers and is not curved. You receive a Pass or Fail result along with your scaled score.
The SPI exam covers ultrasound physics and instrumentation: sound-wave propagation, attenuation, and acoustic impedance; transducers and beam formation; image optimization and resolution; the full range of Doppler concepts (the heaviest domain at 34%); image artifacts; and clinical safety, bioeffects (MI, TI, ALARA), infection control, and quality assurance.
Apply Doppler Concepts is the single largest domain at 34% of the exam — roughly 37 questions. It covers the Doppler effect and shift, Doppler angle, continuous-wave and pulsed-wave Doppler, color and power Doppler, aliasing and the Nyquist limit, wall filters, and spectral-waveform interpretation. Mastering Doppler is the highest-yield use of your study time.
ALARA means As Low As Reasonably Achievable — using the lowest acoustic output power and shortest scan time that still yield a diagnostic image, to minimize potential bioeffects. It anchors the Clinical Safety domain along with the mechanical index (cavitation risk) and thermal index (heating risk) shown on screen under the Output Display Standard.
The SPI exam is taken by sonography students and professionals pursuing ARDMS credentials. Passing SPI plus a specialty examination (for example, Abdomen, OB/GYN, or Adult Echocardiography) earns a registered credential such as RDMS or RDCS. SPI is the physics-and-instrumentation half required for all ARDMS specialties.
Work through the five content domains in order — physics fundamentals, transducers, image optimization, Doppler, and safety/QA. After each module take the checkpoint quiz to find gaps, then drill that domain with our free practice questions and flashcards. Spend the most time on Doppler and image optimization, the two largest domains.
Yes — the full guide, the checkpoints, the glossary, the practice questions, and the flashcards are 100% free, with no account required.
References
- 1.ARDMS. “Sonography Principles & Instrumentation (SPI) Examination.” American Registry for Diagnostic Medical Sonography. ↑
- 2.ARDMS. “SPI Examination Content Outline (V24.1).” American Registry for Diagnostic Medical Sonography. ↑
- 3.AIUM. “As Low As Reasonably Achievable (ALARA) Principle.” American Institute of Ultrasound in Medicine. ↑
- 4.AIUM. “Statement on Biological Effects of Ultrasound In Vivo.” American Institute of Ultrasound in Medicine. ↑
- 5.AIUM. “Prudent Use and Safety of Diagnostic Ultrasound in Pregnancy.” American Institute of Ultrasound in Medicine. ↑
- 6.CDC. “Disinfection & Sterilization Guidelines.” Centers for Disease Control and Prevention. ↑
Sources for the concept answers
Every answer in the SPI concept questions above is drawn from an official primary source:

Career Employer
Career Employer is the ultimate resource to help you get started working the job of your dreams. We cover topics from general career information, career searching, exam preparation with free study materials, career interviewing, and becoming successful in your career of choice.
All PostsCareer Employer’s Editorial Process
Here at Career Employer, we focus a lot on providing factually accurate information that is always up to date. We strive to provide correct information using strict editorial processes, article editing, and fact-checking for all of the information found on our website. We only utilize trustworthy and relevant resources. To find out more, make sure to read our full editorial process page here.
