Career Employer

FREE CIS Study Guide 2026: A Complete HSPA Instrument Specialist Walkthrough

The highest-yield content the HSPA CIS tests — an interactive surgical instrument study guide with built-in flashcards, aligned to the official CIS exam content outline.

Check sections to boost your score

Don't know where to start?

To find us again, just search “Career Employer CIS

By

This free CIS study guide walks through the highest-yield content the exam tests, organized by the six content areas of the official HSPA CIS exam content outline (revised October 2024) — Instrumentation Identification; Inspection, Testing, Integrity & Assembly; Quality & Information Systems; Decontamination Processes; Preparation & Packaging; and Disinfection & Sterilization Processes.[1]

The CIS is an advanced specialty for the technician who owns the instruments: it goes far deeper on identifying, inspecting, testing, repairing, and assembling surgical instruments than any other HSPA exam. This guide is built to teach, not just describe — every content area has the real instrument science, worked exam-style scenarios, comparison tables, labeled diagrams, and built-in flashcards, taught the way the CIS is actually tested.

Read it content area by content area, then round out your prep with our practice questions and flashcards. The CIS builds on the CRCST (the prerequisite) and is a sibling to the CER endoscope-reprocessor credential — this guide stays focused on the surgical-instrument specialty that defines the CIS.

CIS Exam Snapshot

HSPA CIS exam at a glance (2026)
DetailCIS exam
Questions150 multiple-choice
Time limit3 hours (180 minutes)
DeliveryComputer-based at Prometric testing centers
ScoringPass/Fail (section-weighted scaled score; weak-area report provided)
PrerequisiteCurrent, full CRCST certification required first
Experience200 documented hands-on hours (within the past 5 years)
Exam fee~140(datedanchorverifyonmyhspa.org);140 (dated anchor — verify on myhspa.org); 140 retake
RecertificationAnnual; technical CE credits; one renewal fee covers all HSPA credentials
CredentialCertified Instrument Specialist (CIS), awarded by HSPA

Instrumentation Identification alone is 32% of the exam — nearly a third, the single largest area by far.[1] Inspection, Testing, Integrity & Assembly and Quality & Information Systems are each 20%, so those three areas together are about 72% of your score. Weight your study toward instrument identification first, then inspection/testing.

HSPA CIS weighting by content area (Oct 2024 outline)
Instrumentation Identification32% · 32%
Inspection, Testing, Integrity & Assembly20% · 20%
Quality & Information Systems20% · 20%
Decontamination Processes12% · 12%
Preparation & Packaging10% · 10%
Disinfection & Sterilization Processes6% · 6%

Percentages are each area’s share of the scored items, from the official October-2024 outline.[1] This guide teaches all six areas as six study modules. We lead with Instrumentation Identification because it is the largest area and the defining skill of an instrument specialist; the official outline numbers the sections in a different order, which we note in each module.

How the CIS Exam Is Built

The CIS follows the HSPA exam content outline, built from a 2024 job task analysis of working instrument specialists, which groups every scored item into six content areas. This guide teaches all six as study modules, led by Instrumentation Identification — the largest area and the heart of the specialty.[1]

  • Instrumentation Identification (32%) — naming and distinguishing scissors, hemostats, tissue forceps, needle holders, retractors, and rongeurs; specialty and implantable instruments; surgical-grade vs floor-grade; and metals and finishes.
  • Inspection, Testing, Integrity & Assembly (20%) — visual and integrity inspection, inspection and testing tools, rejection criteria, insulation and sharpness testing, preventative maintenance and repair, and set assembly.
  • Quality & Information Systems (20%) — instrument tracking systems and UDI, audits, root cause analysis and process improvement, loaned instrumentation, training and documentation, and water-quality monitoring.
  • Decontamination Processes (12%) — cleaning quality assurance, the cleaning toolbox (water, chemistry, PPE, implements), and specialized handling (ophthalmic, robotics, prions, TASS).
  • Preparation & Packaging (10%) — packaging systems and accessories, material compatibility with sterilization methods, and routine maintenance and quality testing of sealers and rigid containers.
  • Disinfection & Sterilization Processes (6%) — applying sterilization and HLD standards and parameters, and troubleshooting failures, recall, and corrective action.

Instrumentation Identification

Instrumentation Identification is 32% of the exam — the single largest area and the defining skill of the instrument specialist (it is Section II of the official outline).[1] You must be able to name an instrument on sight, place it in the right surgical specialty, and know what it is made of and how that affects its care.

The Instrument Families

Most surgical instruments fall into a handful of families, and the fastest way to identify one is to read its structure before its name. Count the handles and look for a locking : ringed handles with a ratchet mean a or a ; a tweezer-style spring with no ratchet means a . Then read the jaws to name the exact instrument.

Inside each family, small features are the discriminators. vs scissors are told apart by blade shape (heavy and short vs long and thin), not length.

Hemostats are separated by their serration pattern — Mosquito (full, fine), Kelly (serrations only partway down the jaw), and Crile (full serrations). Needle holders differ by their jaws: a Mayo-Hegar has plain cross-hatched jaws; a Heaney has a grooved jaw and a curved handle.

Telling the look-alikes apart
PairThe discriminator
Mayo vs Metzenbaum scissorsBlade shape: Mayo short/thick (tough tissue, suture); Metzenbaum long/thin (delicate)
Kelly vs Mosquito forcepsSerration pattern: Kelly serrated halfway down the jaw; Mosquito fully serrated and finer
DeBakey vs toothed tissue forcepsTip: DeBakey atraumatic fine rows of teeth; toothed (rat-tooth) has sharp interlocking teeth
Mayo-Hegar vs Heaney needle holderJaw/handle: Mayo-Hegar plain straight jaws; Heaney grooved jaw + curved handle (OB/GYN)
Richardson vs Army-Navy retractorRichardson = deep curved blade (handheld); Army-Navy = double-ended flat blades
Kerrison vs pituitary rongeurKerrison = forward-cutting (spine); pituitary = cup jaws for soft tissue/disc

Specialty & Implantable Instruments

Beyond the general families, the CIS tests specialty instrumentation — orthopedic (osteotomes, rongeurs, drills, reamers), neurosurgery (micro-instruments, Kerrison rongeurs), ophthalmic (tiny, delicate, often titanium), cardiovascular (DeBakey forceps, Satinsky clamps), and robotic instruments. Each specialty has its own risk factors: bone cement for orthopedic and dental sets, magnetization of delicate micro-instruments, and residual viscoelastic in ophthalmic surgery (a risk).[1]

Implantable devices are handled with extra rigor. The specialist knows implants by specialty, dimension, and orientation, and follows best practices for inventory, the correct washer cycle, tray assembly, and — critically — sterilization documentation and patient tracking, because an implant is tied to a specific patient and must be quarantined until its is read.[5]

Manufacturing, Metals & Finishes

An instrument’s composition drives its function and its care. Most instruments are stainless steel, protected by (a chromium-oxide layer that resists corrosion). inserts — flagged by gold ring handles — make needle-holder and scissor jaws harder-wearing.

Titanium is light, strong, and non-magnetic for micro and ophthalmic work and is often (colored). Knowing the metal tells you which detergents and chemistries are safe and what failures to watch for.[1]

Marking, Catalogs & Single-Use vs Reposable

Instruments are identified and tracked through marking (laser etching, mechanical etching, color tape, or dipping — applied so it does not harbor soil or interfere with function) and through electronic and manual catalogs used for identification and cross-referencing. The specialist also distinguishes , , and reusable devices, because each is processed and tracked differently — a single-use device must not be reprocessed, while a reposable (e.g., a bipolar cord or breast sizer) is tracked to its defined use count.[4]

Checkpoint · Instrumentation Identification

Question 1 of 10

When identifying surgical instruments, which feature is MOST critical in distinguishing between a Mayo and Metzenbaum scissors?

Inspection, Testing, Integrity & Assembly

Inspection, Testing, Integrity & Assembly is 20% of the exam (Section III of the official outline).[1] This is the specialist’s quality gate: confirm each instrument is clean, verify it works, test it objectively, route the failures for repair, and assemble a complete, correct set.

Visual & Integrity Inspection

Every instrument is inspected under good light and magnification for two things: cleanliness (no soil, , adhesive, or staining) and integrity and function (sharpness, alignment of jaws and tips, working ratchets and , and no cracks, pitting, corrosion, or chips). The first rule still governs: if an instrument is not visibly clean it cannot be sterilized and goes back to decontamination — no later step can fix a cleaning miss.[2]

Testing Tools & Rejection Criteria

Beyond looking, the specialist tests with the right tool and a clear rejection criterion. A inspects inside lumens and channels; an checks electrosurgical coatings for breaks that could burn tissue; a cuts scissors through a size-appropriate material (gauze, index card, dowel rod, leather, or rubber band) to the tip; and a verifies powered and cabled instruments conduct. Each test has a frequency — insulated instruments and scissor sharpness are checked routinely.[1]

Preventative Maintenance & Repair

A dull, damaged, or out-of-spec instrument is not assembled — it is tagged and routed for preventative maintenance (PM) or repair per facility procedure (a sharpened scissor, a re-tipped osteotome, a replaced carbide insert). The specialist also performs PM on the inspection and testing tools themselves — the borescope, the continuity tester, the insulation tester — so the tools that judge the instruments are themselves reliable. Instruments are cleaned and inspected both before and after PM/repair so a vendor return is verified clean before it re-enters service.[1]

Set Assembly & Indicators

Sets are built to a count sheet, with even weight distribution and within weight limits. Hinged instruments are placed open and unlocked so the sterilant reaches every surface, including inside the box lock; delicate tips get protectors, and stringers or racks keep instruments organized. An internal is placed where the sterilant is hardest to reach (typically the center of the set), and an external indicator goes on the outside of every package.[3]

Checkpoint · Inspection, Testing, Integrity & Assembly

Question 1 of 10

When inspecting microsurgical instruments for wear and damage, which of the following is the MOST critical aspect to assess for functionality?

Decontamination Processes

Decontamination Processes is 12% of the exam (Section I of the official outline).[1] The CIS view of decontamination is instrument-focused: the cleaning toolbox, the quality assurance that proves cleaning worked, and the specialized handling of difficult instruments.

Cleaning Science & the Toolbox

Cleaning must always precede disinfection and sterilization — soil shields microbes, and soil left to dry forms that resists removal. The cleaning “toolbox” combines the right water, the correct chemistry (a detergent or enzyme matched to the soil and the instrument, plus adhesive remover where needed), PPE, and the right implements — brushes sized to the lumen,, magnifying devices, and automated mechanical systems. The uses to reach soil in hinges, box locks, and serrations after gross soil is removed.[1]

Cleaning quality assuranceis the specialist’s job: plan and prioritize the workload, run cleaning-verification tests (for example, protein or ATP residual tests on instruments and on the equipment), and identify gaps in the process before they reach a patient.

Water Quality & PPE

Water selection matters at every stage. (tap) is fine for early flushing, but the final rinse uses (purified by RO/DI) so minerals and contaminants are not deposited — poor water causes pitting, staining, and rust on stainless instruments.

defines the water categories and parameters (pH, conductivity, endotoxins). In the decontamination area, full PPE — fluid-resistant gown, gloves, and face/eye protection — is required, and the chemicals are donned, doffed, and stored per their SDS and IFU.[3]

Specialized & High-Risk Instruments

Some instruments need a specialized process: ophthalmic instruments (meticulous cleaning and copious rinsing to prevent ), robotic instruments and powered equipment (specific disassembly and washer cycles per IFU), and flexible endoscopes (which overlap with the CER scope). High-risk pathogens demand special handling: prions () resist normal sterilization and require enhanced processing, and chemotherapy, Ebola, and similar exposures follow specific protocols.[2]

Checkpoint · Decontamination Processes

Question 1 of 10

When selecting a cleaning agent for surgical instruments, which of the following properties is MOST critical to ensure the agent's effectiveness without damaging the instruments?

Preparation & Packaging

Preparation & Packaging is 10% of the exam (Section IV of the official outline).[1] The package is a sterile barrier: it must let the sterilant in during the cycle and keep microbes out afterward, and it must be matched to the sterilization method.

Packaging Systems & Compatibility

The main packaging systems are sterilization wrap (woven or non-woven, square or envelope fold), (paper-and-plastic for small light items), and (reusable boxes with filters, gaskets, valves, and latches). The choice depends on the item and the technique, and the material must be compatible with the sterilization method: cellulose (paper, woven wrap, cotton) works for steam but is not allowed in hydrogen peroxide or gas-plasma cycles, which use instead.[1]

Packaging systems and what to watch for
SystemUseWatch for
Sterilization wrapTrays and sets; square-fold or envelope-foldHoles/tears; method-compatible material (no cellulose in low-temp)
Peel pouchSmall, lightweight single items; see-throughWrite only on the plastic side; seal fully; don't overfill
Rigid containerReusable; protects instrumentsIntact filters, gaskets, valves, and latches each use
Tyvek/plastic (low-temp)Hydrogen peroxide / gas plasma loadsNon-cellulose only; verify the cycle's material compatibility

Drying & Indicator Placement

Instruments are thoroughly dried before packaging: residual moisture dilutes sterilants, blocks penetration, causes wet packs, and in low-temperature methods can abort the cycle. Lumens are dried internally. An internal is placed in the hardest-to-reach spot, and an external indicator on every package distinguishes processed from unprocessed.[3]

Sealer & Container Quality Testing

The packaging equipment is tested too. Heat sealers are checked routinely (seal integrity, temperature, and a seal-strength or dye-penetration test) so pouches do not fail, and rigid containers are inspected each use and maintained per IFU (filters, gaskets, latch function). A weak seal or a missing filter is a sterile-barrier failure even if the cycle ran perfectly.[1]

Checkpoint · Preparation & Packaging

Question 1 of 10

In the sterilization preparation process, which of the following is the MOST important reason for drying instruments after cleaning and before sterilization?

Disinfection & Sterilization Processes

Disinfection & Sterilization Processes is 6% of the exam (Section V, the smallest area).[1] For the CIS it is mostly about applying the right standards and parameters and troubleshooting failures — the in-depth sterilization theory lives in the CRCST.

Standards & Parameters

The specialist applies sterilization and disinfection standards from AORN and AAMI. The intended use sets the process via the : instruments (which enter sterile tissue — most surgical instruments) must be sterilized; semicritical items get at minimum. For HLD, the critical parameters are minimum effective concentration (MEC), contact time, and temperature; for sterilization, the method’s parameters and material compatibility (e.g., Tyvek for low-temperature methods) must be met.[5]

Failures, Recall & Corrective Action

When a cycle fails — a packaging error, a sterilizer fault, a failure code, poor loading, a wet pack, or a positive (failed) — the load is not sterile. The corrective action is a recall: pull every load back to the last negative BI, quarantine those items, remove the sterilizer from service, document, notify, investigate, and reprocess. The makes the recall possible.[3]

Checkpoint · Disinfection & Sterilization Processes

Question 1 of 10

When determining the appropriate disinfection level for surgical instruments, which factor is MOST critical in selecting between high-level disinfection and sterilization?

Quality & Information Systems

Quality & Information Systems is 20% of the exam (Section VI, tied for second-largest).[1] It is the management and data layer of the specialty: tracking instruments, running quality programs, handling loaner sets, training staff, and monitoring water quality.

Tracking Systems & UDI

Instrument-tracking systems (manual or, increasingly, electronic) record every instrument and tray, manage count sheets, and store repair and maintenance histories. They depend on accurate database management and the — the FDA-required code that identifies a specific device through its life. The specialist keeps the data current (system updates, count-sheet maintenance) and understands the systems’ features and challenges, including software updates and data integrity.[4]

Audits, RCA & Process Improvement

Quality work means conducting audits (random sampling, rounding, customer surveys, storage checks) and finding and fixing errors. When something goes wrong, the specialist uses and tools such as cause-and-effect (fishbone) diagrams and to fix the underlying cause, not just the symptom. Process improvement runs on frameworks like Lean, Six Sigma, and SWOT analysis, with key performance indicators (KPIs) and action plans — including reducing instrument turnover and through OR scheduling and inventory.[1]

Loaned Instrumentation

is a tested, high-stakes process. The sets are received and inventoried with their count sheets and IFUs, treated as contaminated and fully decontaminated, inspected, assembled, and sterilized on arrival — never assumed clean or sterile.

After the case, point-of-use treatment, decontamination, restock, and vendor pickup close the loop. Clear SPD–OR–vendor communication and on-time delivery prevent the delays that force IUSS.[1]

Water-Quality Monitoring

The specialist monitors water quality across its categories (utility, critical, steam) and parameters (pH, alkalinity, conductivity, endotoxins), with a defined testing frequency at the point of use — a daily visual check and, for example, a monthly endotoxin test. The right water is used at each stage (final rinse, reverse osmosis, deionized), because water quality directly causes pitting, staining, and rusting on instruments when it is wrong.[3]

Checkpoint · Quality & Information Systems

Question 1 of 10

Which human factor is MOST critical in minimizing the risk of contamination during the handling and transportation of sterile instruments?

How to Use This Study Guide

Work through the guide one content area at a time. After each area, check it off in the contents to raise your exam-readiness score, then drill the same content in our free practice questions and flashcards — active recall and timed practice are what move knowledge into exam-day performance. Because the CIS builds on the CRCST, lean on what you already know and spend your new study time on the instrument-specialist depth.

A high-yield CIS study sequence
  1. 1

    Step 1

    Master Instrumentation Identification (32%): the families, the look-alike discriminators (Mayo vs Metzenbaum, Kelly vs Mosquito), specialty/implant instruments, and metals/finishes. This alone is nearly a third of the exam.

  2. 2

    Step 2

    Lock in Inspection, Testing & Assembly (20%): which tool tests which instrument, the rejection criteria, insulation and sharpness testing, and assembling sets open/unlocked with indicators.

  3. 3

    Step 3

    Work Quality & Information Systems (20%): tracking and UDI, audits/RCA/FMEA, process improvement, the loaner process, and water-quality monitoring.

  4. 4

    Step 4

    Refresh Decontamination (12%) and Preparation & Packaging (10%): the cleaning toolbox, water quality, specialized handling (prions/TASS), packaging-method compatibility, and equipment testing.

  5. 5

    Step 5

    Review Disinfection & Sterilization (6%): standards/parameters and failure/recall — then take full practice tests and aim for 80%+ before exam day.

  • Front-load instrument identification. At 32%, it is the highest-leverage study you can do. Drill names and discriminators until they are automatic.
  • Pair each instrument with its inspection/test. Knowing what a Heaney needle holder is matters more when you also know how to test its ratchet and jaw alignment.
  • Use what you know from the CRCST. Decontamination, packaging, and sterilization overlap with your prerequisite — confirm the instrument-specialist angle and move on.
  • Memorize the high-yield traps. Gold handles = carbide, cellulose is banned in low-temp cycles, loaner sets are never assumed clean, prions need enhanced processing, and only the BI proves sterilization.
  • Then prove it. When a content area feels easy, confirm it with our practice questions and flashcards.

Common questions CIS candidates search and get asked — each answered briefly and backed by an official source (HSPA, CDC, AAMI, FDA, or AORN). Tap any card to test yourself.

CIS Concept Questions

CIS Glossary

Key surgical-instrument and sterile-processing terms in one place. Hover any dotted term throughout the guide for its definition; the full list is below.

CIS
Certified Instrument Specialist — the HSPA specialty credential for a technician who specializes in surgical instrument identification, inspection, testing, repair, and assembly.
HSPA
Healthcare Sterile Processing Association — the certifying body for the CIS, CRCST, and CER, formerly IAHCSMM (renamed January 1, 2022).
CRCST
Certified Registered Central Service Technician — HSPA's entry-level sterile processing credential and the prerequisite for the CIS.
instrument specialist
A sterile processing professional whose advanced focus is the identification, inspection, testing, assembly, repair, and tracking of surgical instruments.
Mayo scissors
Heavy surgical scissors with short, thick blades for cutting tough tissue, fascia, and suture.
Metzenbaum scissors
Surgical scissors with long, thin, delicate blades for fine dissection of soft tissue.
hemostatic forceps
Ringed, ratcheted clamps (hemostats) that occlude bleeding vessels — Mosquito, Kelly, Crile — identified by their serration pattern.
tissue forceps
Tweezer-style, spring-handled forceps with no ratchet that grasp tissue — e.g., DeBakey (atraumatic) and Adson.
needle holder
A ratcheted instrument with short, stout, cross-hatched jaws that holds the suture needle — e.g., Mayo-Hegar and Heaney.
retractor
An instrument that holds an incision open to expose the surgical site — handheld (Army-Navy, Richardson) or self-retaining (Weitlaner, Balfour).
rongeur
A heavy instrument that bites away bone — e.g., the forward-cutting Kerrison used in spine surgery.
box lock
The hinged joint where the two halves of a ringed instrument (hemostat, needle holder) pivot; a frequent site of retained soil.
ratchet
The locking mechanism on the handles of a hemostat or needle holder that holds the jaws closed at set positions.
tungsten carbide
A very hard material inserted into the jaws or blades of premium instruments; gold ring handles flag carbide inserts.
passivation
A chemical process that builds a protective chromium-oxide layer on stainless steel so it resists corrosion.
anodizing
An electrochemical finish (often colored) applied to titanium and aluminum that adds a protective, identifying layer.
reposable
A hybrid device combining a reusable component with limited-use or disposable parts, each tracked to its own use count.
single-use device
A device labeled for one patient and one procedure that must not be reprocessed unless legally remanufactured.
borescope
A thin optical or video probe used to inspect inside lumens and channels for retained soil, debris, or moisture.
insulation tester
A device that checks the protective coating on electrosurgical instruments for breaks, cracks, or pinholes.
continuity tester
A device that verifies powered and cabled instruments conduct electricity correctly.
sharpness test
A functional check that cuts scissors or osteotomes through a size-appropriate material (gauze, index card, leather) to confirm a clean edge.
bioburden
The number of microorganisms on a contaminated item before reprocessing; any visible bioburden fails inspection.
biofilm
A protective layer of microbes and soil that forms when soil dries on an instrument; it resists cleaning and disinfection.
cavitation
The imploding of microscopic bubbles in an ultrasonic cleaner that dislodges fine soil from hinges, box locks, and lumens.
ultrasonic cleaner
A machine that uses cavitation to remove fine soil after gross soil is removed.
utility water
General tap water used for flushing and the early stages of cleaning.
critical water
Treated, purified water (RO/DI) used for the final rinse to avoid mineral and contaminant deposits.
IFU
Instructions For Use — the manufacturer's validated, FDA-cleared method for reprocessing a specific device; legally binding.
TASS
Toxic Anterior Segment Syndrome — sterile eye inflammation from residues on ophthalmic instruments; prevented by thorough cleaning and rinsing.
CJD
Creutzfeldt-Jakob Disease — a prion disease whose agent resists normal sterilization and requires enhanced processing.
Spaulding classification
A framework that ranks devices as critical, semicritical, or noncritical by infection risk to set the reprocessing level.
critical
A device that enters sterile tissue or the vascular system — must be sterilized (e.g., surgical instruments, implants).
high-level disinfection
Destroys all microorganisms except small numbers of bacterial spores; the minimum for semicritical items.
chemical indicator
A device that changes color when a sterilization condition is met — a process check, not proof of sterility.
biological indicator
A vial/strip of resistant spores that, after incubation, proves a cycle actually killed microorganisms.
peel pouch
A paper-and-plastic sterilization package for small, light items; write only on the plastic side.
Tyvek
A flash-spun polyethylene material used to package items for low-temperature sterilization (it is non-cellulose).
rigid container
A reusable sterilization container whose filters, gaskets, valves, and latches are checked each use.
UDI
Unique Device Identifier — an FDA-required code that identifies and tracks a specific medical device.
lot control number
A number linking every package to a specific sterilizer, load, and date, enabling recall and traceability.
RCA
Root Cause Analysis — a structured method that finds the underlying cause of an error so it can be permanently corrected.
FMEA
Failure Mode and Effects Analysis — a proactive method that identifies how a process could fail and the impact of each failure.
IUSS
Immediate Use Steam Sterilization (formerly 'flash') — for items needed right away; minimized as a quality metric.
loaned instrumentation
Vendor-owned instrument sets that must be cleaned, inspected, and sterilized on arrival per their IFUs — never assumed sterile.
ST79
ANSI/AAMI ST79 — the comprehensive guide to steam sterilization and sterility assurance.
ST108
ANSI/AAMI ST108 — the standard for water quality in the processing of medical devices.

CIS Study Guide FAQ

The HSPA CIS (Certified Instrument Specialist) exam has 150 multiple-choice questions, and you have 3 hours (180 minutes) to complete it. The questions are weighted across six content areas, with Instrumentation Identification carrying the most weight at 32%, followed by Inspection/Testing/Assembly and Quality & Information Systems at 20% each.

References

  1. 1.Healthcare Sterile Processing Association (HSPA). “CIS Exam Content Outline (Revised October 2024) & Certification Overview.” HSPA.
  2. 2.Centers for Disease Control and Prevention (CDC). “Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, updated 2024).” CDC.
  3. 3.Association for the Advancement of Medical Instrumentation (AAMI). “ANSI/AAMI ST79 — Comprehensive Guide to Steam Sterilization; ST108 — Water for the Processing of Medical Devices.” AAMI.
  4. 4.U.S. Food and Drug Administration (FDA). “Reprocessing of Reusable Medical Devices & Unique Device Identification (UDI).” FDA.
  5. 5.Association of periOperative Registered Nurses (AORN). “Guidelines for Perioperative Practice — Instrument Cleaning, Sterilization, and Packaging.” AORN.
  6. 101.Healthcare Sterile Processing Association (HSPA). “CIS Exam Content Outline — Instrumentation Identification.” myhspa.org, accessed 20 June 2026.
  7. 102.Centers for Disease Control and Prevention (CDC). “Guideline for Disinfection and Sterilization — Cleaning and Residues.” cdc.gov, accessed 20 June 2026.
Career Employer

Career Employer is the ultimate resource to help you get started working the job of your dreams. We cover topics from general career information, career searching, exam preparation with free study materials, career interviewing, and becoming successful in your career of choice.

Follow Us:

All Posts

Career Employer’s Editorial Process

Here at Career Employer, we focus a lot on providing factually accurate information that is always up to date. We strive to provide correct information using strict editorial processes, article editing, and fact-checking for all of the information found on our website. We only utilize trustworthy and relevant resources. To find out more, make sure to read our full editorial process page here.